TB 9-4220-215-35
SECTION 1
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Depth Gages (General). Calibration and Repair Requirements Worksheets,
DA Form 3758, and the manufacturers' manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. Model variations are described in appendix.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the physical technique.
2. Forms, Records, and Reports.
Forms, records, and
reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in appendix.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required.
Table 1 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/USM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 1. The accuracies listed in table 1 provide a four-to-one ratio between the
standard and the TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 1. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
PNEUMATIC PRESSURE
Range: 0 to 210 psi
(MIS-30859)
STANDARD
Accuracy: 0.06 psi
2