Table 3. PMG Volts
FIELD AMPS/REG VOLTS Meter

Calibration Procedure For Analyzer Test Set Bendix, Type 60B63-5A Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
.
TB 9-4920-359-24
TEST SELECT switch to A‡GEN CT.
(a)
CT CUR READ switch to up.
(b)
GEN/PMG VOLTS RANGE switch to T1.
(c)
400 a GEN T1 to in position.
(d)
Figure 2. CT TEST CURRENT - equipment setup.
(3) Adjust resistance standard to 3300 ohms.
(4) Set Dc power supply to 28 volts.
(5) Set calibrator to 1 V, 400 Hz and press EX SENS. Set calibrator to OPR.
(6) Set power amplifier OUTPUT VOLTAGE RANGE switch to 200 V.
7