SECTION I. IDENTIFICATION AND DESCRIPTION
Dc Voltage

Calibration Procedure For Test Box Assy Pilot Assist/Nulling Model 70700-20678-041 Page Navigation
  1  2  3    4    5  
img
. . . . . . . . . .
TB 9-4920-362-24
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
POWER SUPPLY DCS40-30EM10.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUTOTRANSFORMER
Range: 115 V ac
Ridge, Model 9020A (9020A)
Accuracy: 1%
MULTIMETER
Range: -10.5 to +10.5 mA
Fluke, Model 8840A/AF05
Accuracy: 1.25%
(AN/GSM-64D)
Range: -15.5 to +28.5 V dc
Accuracy: 0.45%
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in TM 55-4920-414-13&P for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Position controls as listed in (1) through (6) below:
(1) AC POWER switch to OFF.
(2) DC POWER switch to OFF.