(5) Set S9 switch (fig.1) to OUT position.
(7) Connect appropriate resistance standard (NO. 1 through NO. 5) to TO
(8) Measure each resistance standard listed in table 6. MEGOHMS meter will
indicate within limits specified.
Actual
SELECT switch
Connection
standard value
value
positions
TO PROBES terminals
+1 div
(MΩ)
LOZ GND
LOZ TO RESISTOR TO
1.0
-4 div
GND
10
100
1 000
LOZ COMP GND
LOZ COMP TO RESISTOR
1.0
---
TO GROUND
10
100
1000
HIZ GND
HIZ TO RESISTOR TO
1.0
---
GROUND
10
100
1000
LOZ-LOZ COMP
LOZ TO RESISTOR TO
1.0
---
LOZ COMP
10
100
1000
LOZ-HIZ
LOZ TO RESISTOR TO HIZ
1.0
---
HIZ
10
100
1000
LOZ COMP-HIZ
LOZ COMP TO RESISTOR
1.0
---
TO HIZ
10
100
1000
EXT
ACROSS EXTERNAL
1.0
---
RESISTANCE
10
TERMINALS
100
1000
11