TB 9-4920-452-35
(2) Repeat technique in (1) above to obtain a 10, 20, 30, 40, 50, 60, 70, and 80
percent cardinal point on TI. Indications as observed on TI will be within limits specified in
(3) Repeat technique in (1) and (2) above for remaining TI gage to be calibrated.
Indications as observed on TI will be within limits specified in applicable table 3 or 4.
b. Adjustments. No adjustments can be made.
a. Performance Check
Figure 1. Frequency meter - test equipment hookup.
(2) Set frequency counter CHANNEL A 50Ω/1MΩ switch to 1MΩ and CHECK/
COMA/SEP switch to SEP.
5