TB 9-4920-455-35
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage items issued as indicated in 4 above, and are not listed in this calibration procedure.
All peculiar accessories are supplied with the TI.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
1
CALIBRATOR
900 and 1800 rpm at 60 Hz
Chadwick-Helmuth, Model 11
DC POWER SUPPLY
Output voltage: 0 to 30 V
HDL, Model LR-10 (LR-10)
Output current: 0 to 5 amps
Range: 10 to 40 Hz
Fluke, Model PM6681/656
Accuracy: 0.25 Hz
(PM6681/656)
1Supplied
with the Vibrex Balance Kit, Model B4591.
CALIBRATOR CHADWICK-HELMUTH, MODEL 11
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI and
TM 55-4920-402-13&P.
d. Unless otherwise specified, all controls and control settings refer to the TI.