TB 9-4920-457-35
Test instrument parameters
Performance specifications
Altitude
Vertical speed (rate of climb)
Airspeed
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 3 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 3. The accuracies listed in table 3
provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 3. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
PNEUMATIC PRESSURE
Druck, Model DPI 145/R (MIS-
STANDARD
45842)
Airspeed
Range: 17.5 to 655 knots
Accuracy: 0.19% of reading
Altitude
Range: 0 to 55,640 ft
Accuracy: 0.21% of reading
Rate of climb
Range: 400 to 5500 ft/minute
Accuracy: 1.75% reading
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common