TB 9-4931-383-50
(a) VOLTAGE RANGE switch to 100.
(b) Voltage dials to 100.00000.
(c) POWER switch to OPR.
(19) Adjust TI voltage dials for a null indication on null detector.
(20) If TI voltage dials do not indicate between 99.99698 and 100.00302, perform
b(4) and (5) below.
(21) If TI voltage dials are within tolerance but not exactly 100.00000, perform b(6)
below.
(22) Set POWER switch to STDBY/RESET.
(24) Set voltage divider to .1000000.
(25) Set operating voltage source (A5) VOLTAGE RANGE switch to 1000 and voltage
dials to 1000.0000.
(26) Set TI and operating voltage source POWER switches to OPR.
(27) Adjust operating voltage source voltage dials for a null indication on null
detector.
(28) Set TI and operating voltage source POWER switches to STDBY/RESET.
(29) Set voltage divider to .999999X
(30) Set TI VOLTAGE RANGE switch to 1000 and voltage dials to 1000.0000.
(31) Set TI and operating voltage source POWER switches to OPR.
(32) Adjust TI voltage dials for a null indication on null detector.
(33) If TI voltage dials do not indicate between 999.9699 and 1000.0300, perform
b(7) and (8) below.
(34) Set TI and operating voltage source POWER switches to STDBY/RESET.
b. Adjustments
29