Table 4. Output Voltage

Calibration Procedure For Dc Power Supply Fluke Models 410B And 410B/At Page Navigation
  1    2    3    4    5  6
img
. . . .
TB 9-4931-401-24
10. Meter Accuracy
a. Performance Check
(1) Connect equipment as shown in figure 2, Connection B.
(2) Adjust OUTPUT VOLTAGE switches for indication of 9 on TI meter.
If
multimeter does not indicate between 8.70 and 9.30 V dc, perform b below.
(3) Set HIGH VOLTAGE switch to STANDBY RESET and POLARITY switch to
- (negative).
(4) Set HIGH VOLTAGE switch to ON and repeat (2) above.
b. Adjustments
(1) Adjust OUTPUT VOLTAGE switches for indication of 9.00 V on multimeter.
(2) Using insulated screwdriver, adjust R150 accessible through hole in bottom of
chassis (under OUTPUT connector), for indication of 9 on TI meter (R).
11. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
7/(8 Blank)