Table 3. Guildline, Model 9711 Calculations Worksheet
SECTION IV. CALIBRATION PROCESS FOR RUBICON, MODEL 2759

Calibration Procedure For Instrument Shunt, Guildline, Model 9711 And Multirange Instrument Shunt Rubicon, Model 2759 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
TB 9-4931-405-40
SAMPLE CALIBRATION TEST REPORT
(Organization) ______________________________________________________________________________
____________________________________________________________________________________________
REPORT OF CALIBRATION FOR
INSTRUMENT SHUNT
GUILDLINE, MODEL 9711
SERIAL NO
______________________________
SUBMITTED BY:
______________________________
______________________________
______________________________
Test instrument
Nominal
Limits of error
resistance
Plug position
Test current
Measured value
(%)
10
A
0.01
.00001
____________________
10000
100
A
0.01
.0001
____________________
1000
0.01
.001
1.0
mA
____________________
100
0.01
.01
10
mA
____________________
10
0.01
.1
100
mA
____________________
1.0
0.01
1
1.0
A
____________________
0.1
0.01
10
2.0
A
____________________
0.01
0.01
10
10
A
____________________
0.01
0.05
100
20
A
____________________
0.001
0.10
300
20
A
____________________
0.000333
Under the conditions stated, the limits of error are as indicated. This calibration is traceable to and
compatible with the National Institute of Standards (NIST) measurements.
Calibration Report No. __________
______________________________
Calibrating Technician
Temperature: __________
Page 1 of 1