SECTION II. EQUIPMENT REQUIREMENTS

Calibration Procedure For Coaxial Thermal Converters Ballantine, Models 1394, 1395, And 1395A Series; Hewlett-Packard, Models 11049A, 11050A, And 11051A; And Filmohm, Model 7842 Series Page Navigation
1  2    3    4    5    6  
img
. . . . . . .
TB 9-4931-406-40
SECTION I.
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Coaxial Thermal Converters, Ballantine, models 1394, 1395, 1395A and
1395B Series; Hewlett-Packard, models 11049A, 11050A, and 11051A; and Filmohm, model
7842 Series.  The manufacturers' manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in the text and tables.
b. Time and Technique.  The time required for this calibration is approximately
3 hours, using the DC and low frequency technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description. TI parameters and performance applications which pertain
to this calibration are in table 1.
Table 1. Calibration Description
Performance specifications
Test instrument parameters
Calibration
Frequency
accuracy
0.1%
50 Hz
AC comparison: Ballantine, Models 1394 Series and
1395A-0.4-M74-98; and
0.1%
400 Hz
Hewlett-Packard Model 1105A-H07-003
0.1%
20 kHz
0.2%
50 kHz
0.25%
100 kHz
0.5%
1 MHz
0.6%
3 MHz
0.6%
10 MHz
1.5%
30 MHz
4.0%
100 MHz
0.1%
50 Hz
AC comparison: Ballantine, Model 1395;
Hewlett-Packard Models 11049A,
0.1%
400 Hz
11050A, and 11051A; and
0.1%
20 kHz
Filmohm, Model 7842 Series
0.2%
50 kHz
0.25%
100 kHz
0.5%
1 MHz
0.6%
3 MHz
0.6%
10 MHz
2