TB 9-4931-427-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
Model FDM 2100. The manufacturers' manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. The
Tracor Model
527E Frequency Difference Meter
has a front
panel phase meter instead of the cathode ray tube used on the Tracor Model 527A and
Hickok Model FDM 2100. Information shown in parenthesis is for Hickok Model FDM
2100. Other variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports.
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) will follow the designated
adjustment. Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to these calibrations are listed in table 1.
Test instrument parameters
Performance specifications
Input
Frequency: 100 kHz to 5 MHz
Voltage: 0.5 to 10 V rms
Frequency difference meter
Range: -10 to +10 parts in 107, 108, 109, 1010, or 1011
Accuracy: 5% of full scale