TB 9-4931-443-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Range: 50 Hz
Boonton, Model 1121 (1121)
Capability: 0.05% distortion
AUTOTRANSFORMER
Range: 105 to 125 V ac
Ridge, Model 9020A
Accuracy: 1%
(9020A)
Range: 50 Hz to 100 kHz
Fluke, Model PM6681/656
Accuracy: 0.25%
(PM6681/656)
Range: 100 V ac
Fluke, Model 8840A/AF05
Resolution: 20 ppm
(AN/GSM-64D)
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove test instrument from protective cover as required for adjustments.
b. Connect test instrument EXCITER UNIT power cable to test instrument POWER
UNIT receptacle P201 located on rear panel.
c. Connect test instrument POWER UNIT power cable to autotransformer.
d. Connect autotransformer to 115 volt ac source and adjust controls for 115 volt output.