TB 9-4931-488-35
(14) Set OUTPUT LEVEL vernier control(s) fully cw.
(15) Measuring receiver will indicate between 99.5 and 100.5%, if not, repeat (11)
through (14) above while readjusting A26A1R19 (A26A1R34 for option 002) (fig. 3) for best
in-tolerance condition.
(16) Set AM switch to OFF.
(17) Set measuring receiver to measure RF power.
(18) For option 002 only, set DBLR OFFSET A26A4R54 (fig. 2) to mid-range.
(19) Adjust LVL A26A4R1 (fig. 2) for a +13.2 dBm measuring receiver indication (R).
(20) Adjust OUTPUT LEVEL vernier control(s) for a +13.0 dBm measuring receiver
indication. Adjust MET A26A4R12 (fig. 2) for TI meter indication of +3 dB (R).
(21) For option 002, proceed to (22) below. For all other models, repeat a above.
(22) Set RANGE MHz switch to 1024-512 and adjust FREQUENCY TUNE control
for FREQUENCY MHz display indication of 800 MHz.
(23) Adjust DBLR LVL A26A4R2 (fig. 2) for a +13.0 dBm measuring receiver
indication (R).