Pulse Characteristics

Calibration Procedure For Tunnel Diode Pulser Tektroniax, Type 067-0681-01 And Ballantine, Model 61252A Page Navigation
  1    2    3  4
TB 9-4931-498-40
(5) Using standard measurement techniques, measure the pulse risetime,
aberrations, and amplitude. The risetime will be 125 ps or less; aberrations will be less
than 1 percent, and amplitude will be approximately 250 mV.
b. Adjustments. No adjustments can be made.
9. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
5/(6 Blank)