TB 9-4931-498-40
(5) Using standard measurement techniques, measure the pulse risetime,
aberrations, and amplitude. The risetime will be 125 ps or less; aberrations will be less
than 1 percent, and amplitude will be approximately 250 mV.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.