CALIBRATION PROCEDURE FOR OSCILLOSCOPE TEKTRONIX, TYPES 5403D40, 5440, AND R5440
SECTION II. EQUIPMENT REQUIREMENTS

Calibration Procedure For Oscilloscope Tektronix, Types 5403D40, 5440, And R5440 Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . . . . . . . . . . . . .
TB 9-4931-501-24
Paragraph
Page
Horizontal centering ............................................
17
11
Horizontal gain.....................................................
18
12
10 and 5 ns timing................................................
19
12
Readout system ....................................................
20
13
Power supply ........................................................
21
14
Final procedure ....................................................
22
15
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Oscilloscope, Tektronix Types 5403D40, 5440, and R5440.
The
manufacturers' manuals were used as the prime data source in compiling these
instructions.  The above equipment will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters
Performance specifications
Vertical amplifier:
Dc to 90 MHz (w/067-0680-00)
Bandwidth
Dc to 60 MHz (w/calibrated 5A48)
3.9 ns or less (w/067-00680-00) 1
Risetime
5.8 ns or less (w/calibrated 5A48)
6% or less (w/067-0680-00)
Aberrations
3% or less (w/calibrated 5A48)
Within 0.5 division of graticule center
Vertical centering
Horizontal amplifier:
Bandwidth
Dc to 2 MHz
See footnote at end of table.
2