TB 9-4931-503-35
Paragraph
Page
CALIBRATION PROCESS FOR TIME BASE,
TEKTRONIX, TYPE 5B40
Preliminary instructions ..................................
Equipment setup ..............................................
External horizontal gain ..................................
Horizontal time ................................................
Final procedure.................................................
CALIBRATION PROCESS FOR DUAL
TRACE AMPLIFIER, TEKTRONIX, TYPE
5A48
Preliminary instructions ..................................
Equipment setup ..............................................
Vertical gain......................................................
Aberrations and risetime .................................
Final procedure.................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Delaying Time Base, Tektronix, Type 5B42; Time Base, Tektronix, Type
5B40; and Dual Trace Amplifier, Tektronix, Type 5A48. The manufacturers' manuals were
used as the prime data sources in compiling these instructions. The equipment being
calibrated will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a.
Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.