TB 9-4931-503-35
Table 3. Main Sweep Timing (Unmag) - Continued
Oscilloscope calibrator
err
Test instrument
display
MAIN SEC/DIV
MARKER
indications
switch settings
output settings
(%)
.1
s
.1
S/D
3
.2
s
.2
S/D
3
.5
s
.5
S/D
3
1
s
1
S/D
3
2
s
2
S/D
4
5
s
5
S/D
4
(7) Press TI SWP MAG pushbutton.
(8) Set TI MAIN SEC/DIV switch and oscilloscope calibrator MARKER output as
NOTE
If necessary, adjust TI MAIN TRIG LEVEL control for a
stable display.
(9) Adjust TI POSITION control to align 2d marker with 2d vertical graticule line
on oscilloscope.
(10) Adjust oscilloscope calibrator EDIT FIELD knob to align 10th marker with 10th
vertical graticule line on oscilloscope. If oscilloscope calibrator err display indication is not
within limits specified in table 4, perform b below.
(11) Repeat technique of (8) through (10) above for remaining rows in table 4.
Oscilloscope calibrator
Test instrument
err display
MAIN SEC/DIV
MARKER output
indications
switch settings
settings
( %)
10
ns
10
nS/D
5
20
ns
20
nS/D
5
50
ns
50
nS/D
4
4
s
S/D
.1
.1
4
s
S/D
.2
.2
4
s
S/D
.5
.5
4
s
S/D
1
1
4
s
S/D
2
2
4
s
S/D
5
5
4
s
S/D
10
10
4
s
S/D
20
20
4
s
S/D
50
50
8