Amplifier Bandwidth

Calibration Procedure For Differential Comparator, Tektronix Type 5A13N Page Navigation
  4    5    6    7    8  9
img
. . . .
TB 9-4931-505-24
b. Adjustments
(1) Adjust R54 (fig. 1) until multimeter indicates 10.00 Vdc (R).
(2) Adjust R57 (fig. 1) until multimeter indicates 1.00 Vdc (R).
16. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective covers on TI if
necessary.
b. Annotate and affix DA label/form in accordance with TB 750-25.