SECTION III. CALIBRATION PROCESS
Figure 2. Terminal plate adjustment locations - bottom view.

Calibration Proceudre For High Voltage Divider, Hallmark Model Kvd 4 Page Navigation
  1    2  3  4  
img
. . . .
TB 9-4931-521-40
Figure 1. Divider accuracy - equipment setup.
NOTE
Observe  spacing  of  high-voltage
leads
when
making
connections shown in figure 1.
c. Set multimeter to read volts dc and set Guard pushbutton To LO.
8. Divider Accuracy
a. Performance Check
(1) Adjust power supply for a measured output of 10,000 V dc, using voltmeter. If
multimeter does not indicate between 0.999 and 1.001 V dc, perform b (1) below.
(2) Reduce output of dc power supply to minimum.
(3) Remove multimeter positive lead from TI 7.5 V terminal and connect to
0.75 V terminal.
(4) Adjust power supply for a measured output of 10,000 V dc, using voltmeter. If
multimeter does not indicate between 0.0999 and 0.1001 V dc, perform b (2) below.