• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
APPENDIX A. TEST INSTRUMENT IDENTIFICATION Continued
APPENDIX A. TEST INSTRUMENT IDENTIFICATION Continued

Calibration Procedure For Attenuators, Fixed And Variable (10 Mhz To 26.5 Mhz) (General) Page Navigation
  14    15    16    17    18  19  20    21  
img
TB 9-4931-523-24
APPENDIX A
TEST INSTRUMENT IDENTIFICATION - Continued
Insertion loss
Nominal
Frequency
Accuracy
at 0 dB
Model number
Manufacturer
attenuation
range
(dB)
(GHz)1
(dB)
(dB)
8491B10
Hewlett-
10
DC to 18
0.6
Packard
8491B20
Hewlett-
20
DC to 18
Packard
DC to 12.4
0.6
12.4 to 18
1.0
8491B30
Hewlett-
30
DC to 18
1.0
Packard
8491B3
Hewlett-
3
DC to 18
0.3
Packard
8491B6
Hewlett-
6
DC to 18
Packard
DC to 12.4
0.3
12.4 to 18
0.54
8492A003
Hewlett-
3
DC to 18
0.3
Packard
8492A006
Hewlett-
6
DC to 18
Packard
DC to 12.4
0.3
12.4 to 18
0.54
8492A010
Hewlett-
10
DC to 18
0.6
Packard
8492A020
Hewlett-
20
DC to 18
Packard
DC to 12.4
0.6
12.4 to 18
1.0
8492A050
Hewlett-
50
DC to 18
1.5
Packard
8493A30DB
Hewlett-
30
DC to 12.4
1.0
Packard
8493B006
Hewlett-
6
DC to 18
Packard
DC to 12.4
0.3
12.4 to 18
0.4
8493B010
Hewlett-
10
DC to 18
0.6
Packard
8493B020
Hewlett-
20
DC to 18
Packard
DC to 12.4
0.6
12.4 to 18
1.0
8493C10
Hewlett-
10
DC to 26.5
Packard
DC to 18
0.3
18 to 26.5
0.5
8493C20
Hewlett-
20
DC to 26.5
Packard
DC to 18
0.5
18 to 26.5
0.6
See footnotes at end of table.