TB 9-4931-523-24
APPENDIX A
TEST INSTRUMENT IDENTIFICATION - Continued
Insertion loss at
Nominal
Frequency
Accuracy
0 dB
Model number
Manufacturer
attenuation
range
(dB)
(GHz)1
(dB)
(dB)
8494B001
Hewlett-
See 8494B
Packard
8495B001
Hewlett-
0 to 70
DC to 18
0.4 dB +0.07
Packard
dB/GHz
DC to 12.4
3% of setting
12.4 to 18
4% of setting
8495D
Hewlett-
0 to 70
DC to 26.5
0.5 dB +0.13
Packard
dB/GHz
DC to 12.4
3% of setting
12.4 to 18
4% of setting
18 to 26.5
7% of setting
0.6 dB +0.09
8496A001
Hewlett-
0 to 110
DC to 4
0.4 or 1.7%
dB/GHz
Packard
whichever is
greater
8496B
Hewlett-
0 to 110
DC to 18
0.6 dB +0.09
Packard
dB/GHz
DC to 12.4
3% of setting
12.4 to 18
4% of setting
8498A030
Agilent
30
DC to 18
1.0
8420731
See PRD Electronics 1100C
951054-2
Boonton
20
DC to 4
0.5
9918-302
Weinschel
30
DC to 18
1.0
9918-602
Weinschel
60
DC to 18
1.5
9918-32
Weinschel
3
DC to 18
0.3
9918-62
Weinschel
6
DC to 18
0.3
9918-102
Weinschel
10
DC to 18
0.5
9918-202
Weinschel
20
DC to 18
0.5
1 Not
calibrated below 10 MHz using this TB.
2
Prepare test report for this item using receiver system indications for system codes: U04, U06, U10, U11, Ul2, U14, F00
and W05.
3 Not checked below -100 dB.
4 Accuracy of measuring receiver must be considered as well as losses associated with any adapter used.
5 Some attenuation settings (>30 dB) will result in accuracy ratios between standard and TI to be <4:1.
6 This is a flatness specification. The maximum recorded attenuation value minus the minimum recorded attenuation value
for each setting will be less than or equal to this accuracy.
7 Deviation from value measured at 4 GHz.
8 Insertion loss for serial number 2524A44630 and below is (0+0.11+1.39 (frequency in GHz). Insertion loss for serial
number 2524A44630 and above is (0+0.2 + 2.3 (frequency in GHz).
A-16