TB 9-4931-540-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
1 M flatness
0.7 dB
Flatness:
1.5 dB
Flatness:
Noise
Resolution
Average
indication
setting
(dBm)
30 kHz
<-100
10 kHz
<-104
3 kHz
<-108
1 kHz
<-111
300 Hz
<-115
100 Hz
<-122
30 Hz
<-127
10 Hz
<-132
3 Hz
<-137
40 Hz <-123 dBm (1 Hz)
Low frequency response
Frequencies: Pwr line freq., 5 kHz, 100 kHz, 1 MHz, and 10 MHz
Accuracy: <-120 dBm
Spurious responses: <-80 dB
Spurious responses: <-80 dB for 50 input
<-70 dB for 1 M input
Accuracy: 20% of BW settings at 3 dB points
Selectivity (shape factor): <11:1 (60 dB BW/3 dB BW)
Flatness: 0.7 dB
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 or AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provide a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common usage
accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration
procedure.