SECTION III. CALIBRATION PROCESS FOR MODELS DV4107C AND RV722
Table 3. Most Significant Decade Switch Linearity Check

Calibration Procedure For Voltage Divider, Electro Scientific Industries, Model Rv722; And General Resistance, Model Dv4107C Page Navigation
  1    2  3  4    5    6  
img
. . . .
TB 9-4935-214-40
8. Decade Linearity
a. Performance Check
(1) Connect equipment as shown in figure 1, connection A.
Figure 1. Decade linearity - equipment setup.
(2) Turn TI decade dials to .999999X and voltage divider decade dials to
.999999 TEN.
(3) Energize dc generator detector and adjust output for +20 V dc, using multimeter.