TB 9-5120-202-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
manual and Federal Specifications, GGG-W-00686C (GSA-FSS) were used as the prime
data sources in compiling these instructions. The equipment being calibrated will be
referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in
table 2.
b. Time and Technique. The time required for this calibration is approximately .5
hours, using the physical technique.
2. Forms,
Records, and
Reports.
a. Forms, records and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications, which
pertain to this calibration, are listed in table 1.
Test instrument parameters
Performance specifications
Torque wrenches:
Type I, II, and V
Range: 0 to 2000 ft-lb
1
Accuracy: 4% of nominal reading from 20% of FS to FS
Type III
Accuracy: CW:
Same as Type I and II
1
6% of nominal reading from 20% of FS
CCW:
Type IV
Accuracy: 5% of setting or as specified below, whichever is
greater from 20% of FS to FS
Size 1: 1/2 in-oz
Size 2: 1/2 in-lb
Range: 0 to 1600 in-oz and 0 to 100 in-lb.
1No
accuracy requirement from 0 to 20% of scale of wrench.