TB 9-5120-206-35
(l) Press the F4 key on the FORCE/TORQUE INDICATOR until you see the
UNIT soft key. Press the UNIT soft key as necessary to select ft-lb.
(m) Press the SIGNAL pq? ? to select the GROSS mode of operation.
Press the F4 key on the TI indicator until you see the Acal soft key. Press the Acal soft
key as necessary to activate the Acal enunciator.
(n) Connect equipment and repeat (1) through (6) above. If TI meter
indication is not within 3 percent of FORCE/TORQUE INDICATOR indication, repeat
(7) above (R).
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
7