TB 9-5120-207-35
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Holding plate, APN 13335453; and socket wrench adapter, 1/2-inch to 3/4-inch,
PN 11655788-3.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
LOAD CELL INDICATOR
Range: 0 to 12,000 counts
BLH, Model 8200B (MIS-23155)
Accuracy: 0.5% of indication
Lebow Associates, Model 2133-126
TORQUE CELL
Range: 0 to 306 ft-lbs
(MIS-26485, Type 1, CL4)
Accuracy: 0.5% of applied torque
(.5% of 20% FS below 20% FS)
SECTION III
6. Preliminary Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
a. Verify that TI is clean and free from defects that would impair its operation.
b. Ensure that equipment has been allowed to stabilize at ambient temperature.
c. Install
holding plate, PN
13335453, on
torque calibration fixture.
d. Install threaded extension handle on TI.
3