Figure 1. Spring tension gages 5210-366-4645 - typical view.

Calibration Procedure For Spring Tension Gages, Models 7 Page Navigation
  1    2    3  4
TB 9-5210-201-24
(2) Reading on scale of TI should coincide with weights suspended in (1) above
within 1/2 graduation TI.
(3) Make a check at all cardinal points (numbered division) of TI.
NOTE
For accurate results, measuring lever must be held at right
angle to line with suspended weight.
(4) For bi-directional gages, rotate TI 180 degrees and repeat performance check.
b. Adjustment. No adjustment can be made.
9. Final Procedure
a. Fold handle forward on TI and replace adjustable tension attachment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
5/(6 Blank)