TB 9-5210-201-50
NOTE
For accurate results, measuring lever must be held at right
angle to line with suspended weight.
(4) For bi-directional gages, rotate unit under test 180 and repeat performance
check.
(5) Annotate block 11, AB, on DA Form 2416.
b. Adjustment. No adjustment can be made.
8. Final Procedure. In
accordance with TM
38-750, annotate and
affix calibration DA
Label 80. When the unit under test cannot be adjusted to within tolerance, annotate and
affix DA Form 2417 (red tag).
5