SECTION III. PRELIMINARY OPERATIONS
Table 3. Distortion (Example)

Calibration Procedure For Optical Flats Page Navigation
  1    2  3  4    5  
img
. . . . . . .
TB 9-5210-202-40
SECTION IV
CALIBRATION PROCESS
8. Flatness
a. Performance Check
(1) Place one TI on another and place on base of monochromatic light (fig. 1).
Figure 1. Monochromatic light - equipment setup.
NOTE
A black cloth or paper placed under bottom TI will make the
band more distinctly readable.
(2) Locate thread as shown in figure 2 and pull taut. Read and record deviation
from straight line of fringes under monochromatic light.
Figure 2. Distortion measurement.
NOTE
Bands should be viewed as near a vertical position as possible.
Use monochromatic light only when required.
(3) Rotate the TI (both flats) approximately 90 degrees and repeat (2) above.
4