TB 9-5210-207-50
Table 10. Inside Micrometer, Tubular, Fixed Head (Metric Measure) Permissible Error - Millimeters
Length ()
Range
Movement of Head
50 to 63
13
0.007
63 to 75
13
0.007
75 to 88
13
0.007
88 to 100
13
0.007
100 to 113
13
0.007
113 to 125
13
0.007
125 to 300
25
0.010
(in 25mm increments)
b. Adjustments. Adjust contact points on each extension rod as required, for zero
indication, when fit is the same as when zero setting micrometer head.
a. At the option of the calibrating laboratory, individual items within a set may be
oiled to prevent rust and sealed in polyethylene film. Upon recalibration, those items with
unbroken seals and no evidence of damage may remain sealed until used.
b. Annotate and affix DA label/form in accordance with TB 750-25.