TB 9-5210-208-24
Tie Rod
Tie Rod
Inside
Sleeve
Caliper Jaw
Gage Block
Unassembled view
Assembled view
11. Inside Function Dial and Digital Calipers
a. Performance Check
(1) Close jaws of TI and set the dial calipers for 0 using the bezel ring on the dial.
Pressing the Zero pushbutton zeros the digital calipers. If there is a minimum inside nib
width it will need to be added to reading.
(2) Using the technique in 9 a (2) and(3) above, measure gage blocks at low range,
mid-range, and at full range. TI will be within r0.001 inch for first 6 inches of bean length
and r0.001 inch for each additional 6 inches of beam length. No adjustment can be made.