TB 9-5210-213-40
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Precision Level, Starrett Model 199. The manufacturer's manual and
Commercial Item Description A-A-50685 were used as the prime data sources in compiling
these instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the physical technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Precision level
Range: 0.0015 inch per 12 inches
Sensitivity: 0.0005 inch per 12 inches
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and the TI. Where the four-to-one ratio cannot be met, the four-to-
one accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories required for this calibration are common
usage accessories issued as indicated in paragraph 4 above and are used in this calibration
procedure.