SECTION III. CALIBRATION PROCESS
APPENDIX A. TEST INSTRUMENT LIST

Calibration Procedure For Bore Measurement Devices Intrimiks Page Navigation
  1    2  3  4  
img
. . . .
TB 9-5210-217-24
Table 3. Diameter Indications.
Test instrument
Test Instrument indication (inches)
Gage Ring sizes (inches)
Style A and B
Min
Max
ranges (inches)
0.275
0.2749
0.2751
0.350
0.3499
0.3501
0.275 0.500
0.425
0.4249
0.4251
0.500
0.4999
0.5001
0.600
0.5998
0.6002
0.500 0.800
0.700
0.6998
0.7002
0.800
0.7998
0.8002
1.000
0.9998
1.0002
1.200
1.1998
1.2002
0.800 1.600
1.400
1.3998
1.4002
1.600
1.5998
1.6002
2.000
1.9998
2.0002
2.400
2.3998
2.4002
2.800
2.7998
2.8002
1.600 4.000
3.200
3.1998
3.2002
3.600
3.5998
3.6002
4.000
3.9998
4.0002
5.000
4.9995
5.0005
6.000
5.9995
6.0005
4.000 8.000
7.000
6.9995
7.0005
8.000
7.9995
8.0005
9.000
8.9995
9.0005
10.000
9.9995
10.0005
8.000 12.000
11.000
10.9995
11.0005
12.000
11.9995
12.0005
b. Adjustments
(1) Before TI is removed from the setting ring, back off the set screw and turn the
barrel so that the reading will conform to the exact Gage Ring size.
(2) Retighten the barrel clamp screw.
(3) Repeat paragraph 8 above.
9. Final Procedure
a. Disconnect and secure all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.