TB 9-5220-211-40
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Sine Plate, Type II, Grade A, Class 2, 5-Inch (GGG-B-121B). Specification
GGG-B-121B was used as the prime data source in compiling these instructions. The sine
plate will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the physical technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. No adjustments to be reported.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
5 in. 0.00010 in.
Length
0.000050 in.
Parallelism of rolls
0.00010 in.
Parallelism of working surfaces
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI. Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.