TB 9-5855-1890-35
f. Loosen mounting post clamp slightly.
g.
Increase tension on
thermocouple TC1
(fig. 3) to
produce a
moderate bow in
thermocouple wire by sliding laboratory clamp forward (toward temperature controller)
in mounting post clamp; then tighten mounting post clamp.
h. If necessary, physically position thermocouple TC1 on heater plate to the position
i. Daub a small amount of thermally conductive paste on side of brass housing of
MRT target (fig. 3) housing (on temperature controller) and place tip of thermocouple
TC2 (fig. 3) in the thermally conductive paste.
j.
Repeat technique of f through h
above for
thermocouple TC2
(fig. 3).
k.
Place leads from differential
thermocouple (fig. 2) on
base of TI near exit port.
5