• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
SECTION I. IDENTIFICATION AND DESCRIPTION
Equipment Setup

Calibration Procedure For Tunable Active Filter Krohn-Hite, Model 3940 Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . . . . .
TB 9-5915-214-24
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
CALIBRATOR
Range: 20 Hz to 1 MHz
Fluke, Model 5720A (5720A) (p/o MIS-
35947); w amplifier, Fluke 5725A/AR
Accuracy: r 0.25%
(5725A/AR)
MULTIMETER
Range: 0 to 5 V dc
Fluke, Model 8840A/AF05
(AN/GSM-64D)
Accuracy: r 0.025%
TRUE RMS
Range: -2 to -25.4 dB
Fluke, Model 8922A/AA (8922A/AA)
VOLTMETER
Accuracy: r 0.25%
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in paragraphs 7 through 12 are not within tolerance,
perform Section IV, Adjustment Process. After adjustments are made, repeat paragraphs 7
through 12. Do not perform Section IV if all other parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
the calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Connect TI to a 115 V ac source.
NOTE
When  POWER  rocker  switch  is  pressed  to  ON,  TI
automatically performs a self-test sequence. Upon successful
completion of self-test, TI will be in normal operating mode.
3


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.