2nd Quadriatic Frequency Calibration - Continued

Calibration Procedure For Tunable Active Filter Krohn-Hite, Model 3940 Page Navigation
  11    12    13    14    15  16
img
.
TB 9-5915-214-24
v. Set calibrator and TI frequencies for a 2 MHz output, using technique of c above to
enter TI frequency, but press MEGA instead of KILO. If true rms voltmeter does not
indicate between -2.36 and -3.75 dB, readjust C234 (fig. 1) until true rms voltmeter
indicates between -2.36 and -3.75 dB.
w. Set calibrator and TI frequencies for a 256 kHz output, using technique of c above to
enter TI frequency. Verify true rms voltmeter indicates between -2.76 and -3.29 dB.
x. Repeat technique of a through w above for CH 2.