TB 9-5915-214-24
(2) KILO parameter key.
(3) FREQ parameter key.
c. Connect calibrator to INPUT CH 1 using 50 : termination and adjust calibrator for
a 20 kHz, 0.1 V output.
d. Connect true rms voltmeter to rear panel CH 1 OUTPUT using 50 : termination
e. Adjust calibrator for a 2 MHz, 0.1 V output. If true rms voltmeter does not indicate 0
dB, r0.01 dB, adjust C265 (fig. 1) until true rms voltmeter indicates 0 dB, r0.01 dB, or as close
to 0 db as possible.
f Press corresponding keys as listed in (1) through (5) below and verify CUTOFF
FREQUENCY indicates 1.500.
(1) 1 data entry key.
(2) . data entry key.
(3) 5 data entry key.
(4) MEGA parameter key.
(5) FREQ parameter key.
g. Adjust calibrator for a 5 MHz, 0.1 V output. If true rms voltmeter does not indicate 0
dB, r0.05 dB, adjust R249 (fig. 1) until true rms voltmeter indicates 0 dB, r0.05 dB, or as close
to 0 db as possible.
h. Repeat b through g above until both indications are within tolerance.
i. Repeat technique of a through h above for CH 2.
19. 2nd Quadriatic Frequency Calibration
a. Press MODE parameter key until CUTOFF FREQUENCY display indicates bYP.
b. Press INPUT(db) and OUTPUT GAIN(db) control keys until displays indicate 00.
c. Press corresponding keys as listed in (1) through (5) below and verify CUTOFF
FREQUENCY indicates 175.0.
(1) 1 data entry key.
(2) 7 data entry key.
(3) 5 data entry key.
(4) KILO parameter key.
(5) FREQ parameter key.
d. Connect calibrator to INPUT CH 1 using 50 : termination.
e. Connect true rms voltmeter to OUTPUT CH 1.
f. Adjust calibrator for a 175 kHz, 0.1 V output and reference true rms voltmeter to 0 dB.
g. Press MODE parameter key until CUTOFF FREQUENCY display indicates L.P.