VSWR Check
Final Procedure

Calibration Procedure For Power Standard Assemblies 7913995, 7916259, Model 4098A Option 11 Page Navigation
  4    5    6    7    8  9  10    11  
img
. .
TB 9-5985-313-40
NOTE
In the following checks, max hold should be used on spectrum
analyzer.
(1) Connect equipment as shown in figure 4, connection A.
Figure 4. VSWR check - equipment setup.
(2) Setup spectrum analyzer to make a reference trace, (trace 1), for approximately
1.00 VSWR while sweeping signal generator across frequency range of TI.
(3) Disconnect termination no. 2 from VSWR bridge and connect termination no. 3
as shown in connection B.
(4) Blank 1.00 trace, (trace 1), on spectrum analyzer.  Sweep signal generator
across frequency range of TI and make a reference trace, (trace 2), for 1.20 VSWR.
(5) Disconnect termination no. 3 from VSWR bridge and
connect
equipment
as
shown in connection C.
(6) Blank 1.20 trace, (trace 2), on spectrum analyzer. Sweep signal source across
frequency range of TI and make a trace (trace 3).
(7) Display all 3 traces on spectrum analyzer. Trace 3 will be less than the 1.20
VSWR reference, (trace 2), established in (4) above.
10