TB 9-6625-1014-35
d. Adjust TI mechanical zero indication on meter.
a. Performance Check
(1) Position TI controls as listed in (a) through (d) below:
(a)
FUNCTION switch to RX10.
(b)
GEN-DET switch to BATT-TEST.
(c)
DET GAIN control to PWR OFF.
(d)
LRC controls to 0000.
(2) Connect leads from resistance bridge (part of resistance measurement system),
UNKNOWN 1 terminal to TI EXT D-Q 1 terminal, and from UNKNOWN 2 terminal to TI
RL 1 terminal.
(3) Adjust resistance bridge to obtain the best null indication on dc detector. Record
resistance value.
(4) Set LRC dials to 0.020.
(5) Adjust resistance bridge for null indication on dc detector.
(6) Subtract resistance value recorded in (3) above from value obtained in (5) above;
difference will be between 19 and 21 Ω.
(7) Repeat technique of (4) through (6) above for remaining LRC dial settings listed
in table 5. Resistance bridge will indicate within computed limits specified.
b. Adjustments. No adjustments can be made.
Test instrument
Resistance standard
Computed difference
LRC dials
indications
Min
Max
0.040
40
39
41
0.060
60
59
61
0.080
80
79
81
0.0X (100)
100
99
101
0.100
100
99.95
100.05
0.200
200
199.90
200.10
0.300
300
299.85
300.15
0.400
400
399.80
400.20
0.500
500
499.75
500.25
0.600
600
599.70
600.30
0.700
700
699.65
700.35
0.800
800
799.60
800.40
0.900
900
899.55
900.45
0.X
1000
999.50
1000.50
1.000
1000
999.5
1000.5
2.000
2000
1999.0
2001.0
3.000
3000
2998.5
3001.5
4.000
4000
3998.0
4002.0
5.000
5000
4997.5
5002.5