Table 11. General Radio, Type 1419M (MX-4618/U) Capacitance
APPENDIX A. SAMPLE CALIBRATION TEST REPORT FOR VARIABLE AIR CAPACITOR

Calibration Procedure For Standard Capacitors Fixed And Variable (General) Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. .
TB 9-6625-1062-24
Table 12. ZM-59/U Capacitance - Continued
Test instrument
Capacitance measurement system
Indications (F)
Selector
Capacitance value
(F)
switch setting
Min
Max
2
0.02
0.01990
0.02010
3
0.03
0.02985
0.03015
4
0.04
0.03980
0.04020
5
0.05
0.04975
0.05025
6
0.06
0.05970
0.06030
7
0.07
0.06965
0.07035
8
0.08
0.07960
0.08040
9
0.09
0.08955
0.09045
2
1
0.001
0.000995
0.001005
2
0.002
0.001990
0.002010
3
0.003
0.002985
0.003015
4
0.004
0.003980
0.004020
5
0.005
0.004975
0.005025
6
0.006
0.005970
0.006030
7
0.007
0.006965
0.007035
8
0.008
0.007960
0.008040
9
0.009
0.008955
0.009045
1Turn
outer dial to 0.
2Turn
center dial to 0.
10. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. When all parameters are within tolerance, annotate and affix DA Label 80 (US
Army Calibrated Instrument).  When the TI receives limited or special calibration,
annotate and affix DA Label 163 (US Army Limited or Special Calibration). When the TI
cannot be adjusted within tolerance, repair the TI in accordance with the maintenance
manual. When repair is delayed for any reason or the TI cannot be repaired with local
resources, annotate and affix DA Form 2417 (US Army Calibration System Rejected
Instrument) and inform the owner/user accordingly in accordance with TB 750-25-1.
NOTE
For fixed value, non adjustable capacitors: If the TI has not drifted more
than the allowable tolerance from the previous test report, the TI will be
considered within tolerance.