TB 9-6625-1062-24
Test instrument
Indications (F)
Selector
Capacitance value
(F)
switch setting
Min
Max
2
0.02
0.01990
0.02010
3
0.03
0.02985
0.03015
4
0.04
0.03980
0.04020
5
0.05
0.04975
0.05025
6
0.06
0.05970
0.06030
7
0.07
0.06965
0.07035
8
0.08
0.07960
0.08040
9
0.09
0.08955
0.09045
2
1
0.001
0.000995
0.001005
2
0.002
0.001990
0.002010
3
0.003
0.002985
0.003015
4
0.004
0.003980
0.004020
5
0.005
0.004975
0.005025
6
0.006
0.005970
0.006030
7
0.007
0.006965
0.007035
8
0.008
0.007960
0.008040
9
0.009
0.008955
0.009045
1Turn
outer dial to 0.
2Turn
center dial to 0.
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. When all parameters are within tolerance, annotate and affix DA Label 80 (US
Army Calibrated Instrument). When the TI receives limited or special calibration,
annotate and affix DA Label 163 (US Army Limited or Special Calibration). When the TI
cannot be adjusted within tolerance, repair the TI in accordance with the maintenance
manual. When repair is delayed for any reason or the TI cannot be repaired with local
resources, annotate and affix DA Form 2417 (US Army Calibration System Rejected
Instrument) and inform the owner/user accordingly in accordance with TB 750-25-1.
NOTE
For fixed value, non adjustable capacitors: If the TI has not drifted more
than the allowable tolerance from the previous test report, the TI will be
considered within tolerance.