TB 9-6625-1098-35
NOTE
It may be necessary to reinstall the shorting bar between TI
and chassis ground terminals below 1 kHz.
(f) MODE switch to MANUAL (Hewlett-Packard Models 333A and 334A).
(3) Adjust calibrator frequency for 5 kHz and amplitude for an indication of 3 on TI
meter 0 to 3 scale.
(4) Set FUNCTION switch to DISTORTION.
(5) Adjust BALANCE controls and calibrator frequency for a null indication on TI
meter (METER RANGE switch may be reduced to obtain best null indication). If
calibrator frequency does not indicate between 4.75 and 5.25 kHz, perform b below.
(6) Set METER RANGE switch to .3 VOLTS and FUNCTION switch to
SET LEVEL.
(7) Repeat (2) through (6) above, using frequencies, control settings, and indications
Test instrument
Calibrator
FREQUENCY
frequency
RANGE
FREQUENCY
Nominal
indications
frequency
switch settings
dial settings
Min
Max
20 kHz
X1K
20
19.0 kHz
21.0
kHz
60 kHz
X1K
60
57.0 kHz
63.0
kHz
50 kHz
X10K
5
47.5 kHz
52.5
kHz
200 kHz
X10K
20
190 kHz
210
kHz
600 kHz
X10K
60
540 kHz
660
kHz
6 kHz
X100
60
5700 Hz
6300
Hz
2 kHz
X100
20
1900 Hz
2100
Hz
500 Hz
X100
5
475 Hz
525
Hz
50 Hz
X10
5
47.5 Hz
52.5
Hz
200 Hz
X10
20
190 Hz
210
Hz
600 Hz
X10
60
570 Hz
630
Hz
60 Hz
X1
60
57 Hz
63
Hz
20 Hz
X1
20
19 Hz
21
Hz
b. Adjustments
(1) Position switches as listed in (a) through (c) below:
(a) METER RANGE to .3 VOLTS.
(b) FUNCTION to SET LEVEL.
(c) FREQUENCY RANGE to X1K.
(3) Set FUNCTION switch to DISTORTION.
(4) Adjust FREQUENCY dial for a null indication on TI meter. (Set METER
RANGE switch as necessary for sharp null.)
12