Resistance Accuracy
Table 4. 10 M?/Step V = 100TI

Calibration Procedure For Decade Resistor, Iet Model Hrrs-Q-4-1M-10Kv-Rm Page Navigation
  1    2    3  4  5  
img
. . . . . .
TB 9-6625-115-24
Figure 1. Resistance measurement connection.
(8) TI will indicate within the limits of table 3.
(9) Repeat technique of (2) through (8) for remaining values in table 3.
(10) Set calibrator output to STBY.
(11) Repeat technique of (2) through (10) above for remaining tables, except in
step (3), set calibrator output to 100 V for table 4 and 1000 V for tables 5 and 6.
(12) Reduce calibrator output to 0.00 V dc and set to STBY.
Table 3. 1 MΩ/Step V = 10
TI
Multimeter
TI Resistance
Min (MΩ)
Max (MΩ)
switch position
Indication (Vs)
(Rx)
0-0-0-1
0.99990
1.00010
0-0-0-2
1.99980
2.00020
0-0-0-3
2.99970
3.00030
0-0-0-4
3.99960
4.00040
0-0-0-5
4.99950
5.00050
0-0-0-6
5.99940
6.00060
0-0-0-7
6.99930
7.00070
0-0-0-8
7.99920
8.00080
0-0-0-9
8.99910
9.00090
0-0-0-10
9.99900
10.00100