TB 9-6625-1200-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This provides bulletin instructions for the
manuals were used as the prime data sources in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout this
bulletin.
a. Model Variations. None that effects calibration.
b. Time and Technique. The time required for this calibration is approximately 3
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Power input requirement
115 V ac 10% at 50 to 400 Hz
Vertical sensitivity
Range: 0 to 20 V/div
Accuracy:
3%, X1
4% for channel 2 in X10 GAIN position
( 7.5% for S/N below 20,000)
Calibrator:
Internal
Range: 0.2 V square wave
Accuracy:
1.6% for S/N below 20,000
2.5% for S/N above 20,000
Calibrator jack
Range: 2 V square wave
Accuracy:
3.6% for S/N below 20,000
1.5% for S/N above 20,000
Risetime
24 ns or less with aberrations of 0.15 div or less
Sweep timing
Range: 0.5 s to 0.5 s/div
Accuracy:
3%
5% in X10 MAG