TB 9-6625-1200-35
10. Magnifier
Registration
a. Performance Check
(1) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 output to TI CH
1 input.
(3) Position controls as listed in (a) through (h) below.
(a) CH1-CH2 mode switch to CH1.
(b) CH 1 VOLTS/DIV switch to .5
(c) TRIGGERING source switch to CH1.
(d) TRIGGERING coupling switch to AC.
(e) TIME/DIV switch to 1 mSEC and VARIABLE control to CAL.
(f) TRIGGERING LEVEL control for stable display.
(g) Pull out X10 MAG switch.
(h) Adjust horizontal POSITION control to align first time marker to center
vertical graticule line.
(4) Push in X10 MAG switch. If first marker does not remain aligned behind center
vertical graticule line within 1 minor division, perform b below.
b. Adjustments.
Adjust R535 (fig. 2) until
first marker is
positioned to
center vertical
graticule line for both X10 and X1 MAG switch positions (R).