(8) Vary function/arbitrary generator frequency, from 100 Hz to 50 kHz while
maintaining indication recorded in (7) above. Function/arbitrary generator will indicate
between 8.9 and 9.1 V.
(9) Set GAIN switch to 20 dB.
(10) Repeat (7) and (8) above.
(12) Substitute true rms voltmeter for multimeter.
(13) Repeat (7) and (8) above, while varying function/arbitrary generator frequency
from 10 Hz to 1 MHz. If true rms voltmeter does not indicate between 7.2 and 10.8 V,
perform b below.
(14) Set GAIN switch to 40 dB and repeat (13) above. If out-of-tolerance conditions
exist, set GAIN switch to 20 dB and perform b below.
b. Adjustments
(1) Set function/arbitrary generator frequency for 1 MHz and output for 1 V (verify
with true rms voltmeter).
(2) Adjust C5 1MHz ADJ(fig. 5) until true rms voltmeter indicates 8.5 V (R).
Figure 5. Hewlett-Packard Model 465A - top view.
17. Distortion
a. Performance Check
(1) Connect function/arbitrary generator to audio analyzer.
(2) Adjust function/arbitrary generator frequency for 10 Hz and amplitude for 1 V.