TB 9-6625-1314-35
Table 11. Ac Volts
Test instrument
Ac voltage calibrator
Adjustments
Optimum
RANGE
switch
Indications
value
Adjustments
setting
Indication
Min
Max
(V).
10 V
8.000
1000 Hz
7.9900
8.0100
---
---
10 V
6.000
1000 Hz
5.9920
6.0080
---
---
10 V
4.000
1000 Hz
3.9940
4.0060
---
---
10 V
2.000
1000 Hz
1.9960
2.0040
---
---
10 V
9.000
50 Hz
8.9890
9.0110
---
---
10 V
9.000
20 kHz
8.98901
9.01101
9.0000
A1C1 (R)
10 V
9.000
50 kHz
8.9880
9.0120
---
---
10 V
9.000
100 kHz
8.9680
9.0320
9.000
A1L2 (R)
100 V
90.00
1000 Hz
89.890
90.110
90.000
A2R2 (R)
100 V
90.00
50 Hz
89.890
90.110
---
---
100 V
90.00
20 kHz
89.890 2
90.1102
90.000
A2C7 (R)
100 V
90.00
50 kHz
89.880
90.120
---
---
100 V
90.00
100 kHz
89.680
90.320
---
---
1000 V
900.0
1000 Hz
898.90
901.10
900.00
A2R4 (R)
1000 V
900.0
50 Hz
898.90
901.10
---
---
1000 V
900.0
20 kHz
898.903
901.103
900.00
A2C4 (R)
1000 V
900.0
50 kHz
898.80
901.20
---
---
1000 V
900.0
100 kHz
896.80
903.20
---
---
1C06-3455A
limits are 8.9780 to 9.0220.
2C06-3455A
limits are 89.780 to 90.220.
3C06-3455A limits are 897.80 to 902.20.
Figure 4. Ac/dc range unit - adjustment locations.
23