TB 9-6625-132-24
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals for this TI.
d. When indications specified in paragraphs 8 through 12 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 12. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all control and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective cover as needed to make adjustments.
Replace cover after
completing the adjustments.
CAUTION
Set Power Module to OFF before installing or removing TI.
Damaging voltages to the equipment may be encountered
during the sequence of these test procedures. All necessary
precautions during the conduct of these tests must be observed.
b. Insert TI into Power Module.
c. Connect TI to proper power source. Set Power switch to ON and allow a 20 minute
warm-up period.