Sinewave Distortion - Continued
Final Procedure

Calibration Procedure For Function Generator Model Fg501A Page Navigation
  5    6    7    8    9  10  11  
img
. . .
TB 9-6625-132-24
Figure 1. Adjustment locations.
12. Squarewave and Pulse output
a. Performance Check
(1) Connect TI OUTPUT to oscilloscope channel 1 input using a 50 Ω feedthrough
termination and X10 attenuator.
(2) Set TI controls as indicated in (a) through (g) below.
(a) FREQUENCY Hz
20
105
(b) MULTIPLIER
(c) Waveform pushbutton
(Square)
(d) VAR SYMM
off
(e) FREE RUN
in
(f) ATTENUATOR
0 dB
(g) AMPL (variable)
fully CW
(3) Connect TI TRIG OUTPUT to oscilloscope channel 2 input.
(4) Set oscilloscope channels 1 and 2 for DC input coupling.
(5) Adjust oscilloscope controls to measure both rise time and fall time of TI output
signal. If rise time and fall time are not less than 25 ns, perform b below.
(6) Set TI VAR SYMM to on and adjust VAR SYMM control for a pulse waveform on
oscilloscope.